| Bus Interface | PCI 64-bit / 66 MHz1 |
|---|---|
| Dimensions | Short length, full height, PCI 32-bit, 64-bit, or PCI-X card slot2 |
| Weight | Approximately 0.3 kg |
| Frame Buffer Memory | 256 MB of high-speed memory for fast image caching and loading |
| Display Support | Dual DVI-I connectors supporting 1 or 2 displays (color or grayscale) |
| Digital Communications | DDC-CI (support available for custom EDIDs) |
| Display Calibration | As supplied by each display vendor, or with purchase of third-party hardware/software solutions3 |
| Display Configurations | Grayscale/Grayscale, Grayscale/Color, Color/Grayscale, Color/Color |
| Grayscale Resolution | Matrox MED5mp: From 1.3 to 5 MP up to 2560x2048 (landscape) and up to 2048x2560 (portrait) |
| Matrox MED4mp: 1.3, 2, 3 and 4 MP up to 2560x1600 (landscape) and up to 1600x2560 (portrait)4 | |
| Matrox MED3mp: 1.3, 2 and 3 MP up to 2048x1536 (landscape) and up to 1536x2048 (portrait) | |
| Matrox MED2mp: 1.3 and 2 MP up to 1920x1200 (landscape) and up to 1200x1920 (portrait) | |
| Grayscale Settings | 8-bit input/10-bit output supporting Fixed Linear, Fixed Non-Linear, or Programmable palettes offering up to 256 distinct shades of gray (256 JNDs) |
| 10-bit input/10-bit output with up to 1024 distinct shades of gray5 | |
| Mixed mode 8- or 10-bit grayscale and 32-bit true color | |
| Dual Programmable LUTs | Import a non-linear gamma ramp from a calibration device and independently program either display's Look Up Table outputs |
| OS Support | Microsoft® Windows® XP, Windows 2000, Windows NT 4.0 |
| VGA Support | Built-in controller displays boot messages on either a fixed frequency or multi-sync monitor |
| Software Support | Supports OpenGL® 1.3 and/or DirectDraw/DirectX compatible applications |
| Industry Compliance | DICOM Pt 14 grayscale compliant |
| EMI Certifications | FCC Class A |
| Environmental | Operating temperature from 0 to 55 degrees Celsius |
| Atmospheric | Operating pressure from about 650 to 1013 hPa |
| Humidity | Operating humidity from 20% to 80%, non-condensing |
1 PCI-64/66 is backward compatible with PCI 32-bit/33 MHz
2 PCI card will operate in 3.3- or 5-volts slots
3 See our list of compatible display or choose a 3rd party software solution (e.g. Image Smiths, Inc. or QUBYX).
4 Matrox MED4mp supports only one digital output at 4MP. The second output in analog only at a maximum 1600x1200 (landscape) or 1200x1600 (portrait) resolution.
5 1024 simultaneous shades of gray supported with 10-bit displays